No light from light source |
Power switch off |
Power cord improperly connected or not plugged in |
Problem with power supply |
Blown fuse |
Faulty bulb/defective bulb contacts |
No light seen through microscope objectives |
Light source covered |
Field diaphragm closed |
Aperture/iris/condenser diaphragm closed |
Objective not in the light path |
Field only partially illuminated |
Condenser off-center |
Aperture/iris/condenser diaphragm closed too much for the objective being used |
Field diaphragm partially closed |
Objective only partly in light path |
Electrical cord lying in the field |
For Nikon 50i: neutral density filter partly pushed in |
For Nikon 50i: optical path switching lever (for camera) in incorrect position |
Weak illumination |
Light adjustment knob/rheostat/lamp voltage turned too low |
Light source covered or burning out |
Aperture/iris/condenser diaphragm closed too much for the objective being used |
Condenser too low or not centered |
Using oil immersion lens with no oil |
For Nikon 50i: neutral density filter engaged |
For Nikon 50i: preset switch has been engaged (only required for camera use) |
Uneven illumination |
Light source not centered |
Condenser not centered |
Field of view too bright |
Light adjustment knob set too high |
Image cannot be focused |
Slide very dirty/greasy |
Slide up-side down |
All objectives dirty |
Dirty eyepieces |
Diopter adjustment improperly set |
Vibration on the bench |
Image very yellow |
Blue filter missing |
Light adjustment knob set too low |